CellProfiler quantitative microscopy
Use this skill when a user needs a repeatable CellProfiler .cppipe, nuclear counts, nuclear
fluorescence, or batch microscopy measurements. The bundled assay accepts one 2D grayscale
TIFF nuclear channel per field, with bright nuclei on a dark background. For volumetric
segmentation, multichannel cell painting, or tissue-specific models, design a separate pipeline
and validate those assumptions rather than silently projecting or splitting the images.
工作流程
- Establish the acquisition unit: plate, well, site, time point if present, pixel size, nuclear channel identity, camera bit depth, exposure, and biological replicate. Keep original image intensities. Convert proprietary formats explicitly with Bio-Formats before using this helper.
- Create the CSV manifest below.
image_pathis absolute or relative to the manifest; sample IDs use letters, digits, dots, dashes, or underscores; sample IDs and plate/well/site combinations are unique. TIFFs must be uint8 or uint16, single plane, and nonconstant. The helper rejects RGB, z-stacks, and float images rather than guessing channels. - Use assets/nuclei.cppipe as a starting pipeline: LoadData → IdentifyPrimaryObjects → intensity/size measurements → outline overlay → CSV export. The initial diameter range is 8–80 pixels, with global Otsu thresholding, no threshold smoothing, and border objects excluded. Calibrate this range from representative images and acquisition pixel size before comparing conditions.
- Run a small pilot spanning controls, low/high density, dim images, and plate edges. Inspect
saved overlays for missed nuclei, splits, merges, and edge exclusions. Adjust thresholding
and declumping in CellProfiler, export the tuned
.cppipe, and pass--pipelineto preserve it. Do not choose settings separately for each treatment to make their counts agree. - Freeze the tuned pipeline and analyze the batch. Review input saturation warnings, zero counts, count/area distributions, and control behavior. Aggregation for inference belongs at the biological replicate level; thousands of cells from one well are not independent wells.
Run the bounded assay
From this skill directory, create images.csv:
sample_id,image_path,plate,well,site
control_A01_1,images/control_A01_1_DAPI.tif,Plate1,A01,1
python scripts/nuclei_assay.py prepare images.csv load_data.csv
python scripts/nuclei_assay.py run images.csv results --executable cellprofiler
python scripts/nuclei_assay.py summarize results
run requires a fresh/empty output directory and executes CellProfiler with -c -r, an explicit
pipeline, --data-file, and output folder. It records the command, pipeline checksum, input image
checksums, and sample QC in assay_qc.json; CellProfiler output goes to cellprofiler.log.
A failed process stays failed, with its log available for diagnosis. Rerun in a new output folder.
The executable can also be the CellProfiler application launcher or a local container launcher;
see references/runtime-and-qc.md for the tested container,
filesystem mapping, and verification evidence. prepare and summarize work without CellProfiler.
Interpret the outputs
Image.csv: one image/field row, includingCount_Nucleiand acquisition metadata.Nuclei.csv: one accepted object per row, with mean/integrated DNA intensity, area, and shape.*_nuclei.png: green nuclear boundaries over the input image for visual QC.assay_qc.json: count consistency and finite normalized intensity checks, plus saturation flags.
LoadData ignores camera metadata for scaling in this asset and divides by the integer storage maximum: uint8 → 255, uint16 → 65535. A 12-bit camera stored in uint16 therefore has a maximum near 0.0625. Do not compare intensities across different bit depths, exposures, gains, or staining batches without an explicit calibration. A saturated image can pass segmentation while its intensity measurement is unusable. Illumination correction and background subtraction are assay-specific additions; this starter does neither.
A count check cannot prove correct segmentation. Inspect overlays and independently annotated fields; report boundary exclusions and segmentation errors alongside the biological result. The synthetic integration test validates a known three-nucleus example, not assay performance on unseen cell types.